Silicon steel sheet iron loss meter measuring principle and use precautions
Release time:
2020-05-14
Source:
Www.fe2100.com Hunan Yongyi Technology Co., Ltd.
The silicon steel sheet iron loss meter is a special instrument for measuring the iron loss of silicon steel sheet. It can be displayed directly through the digital meter head and can intuitively understand the loss of electrical steel sheet. The instrument is mainly used for on-site detection of iron loss value of single silicon steel sheet. Can be widely used in silicon steel sheet production, trade, motors, transformers and other applications of silicon steel sheet manufacturing products industry for silicon steel sheet quality management
Silicon steel sheet iron loss meterIt is a special instrument for measuring the iron loss of silicon steel sheet. It can be displayed directly through the digital meter head and can intuitively understand the loss of electrical steel sheet. The instrument is mainly used for on-site detection of iron loss value of single silicon steel sheet. Can be widely used in silicon steel sheet production, trade, motors, transformers and other applications of silicon steel sheet manufacturing products industry for silicon steel sheet quality management. FE-30SST LOSSMETER silicon steel sheet iron loss measuring instrument is a closed-loop control measuring instrument designed with a combination of single-chip technology and analog electronic technology. Silicon steel sheet iron loss meter has the characteristics of small size, light weight, easy to carry, simple and fast operation, convenient sampling and so on. FE-30SST
Silicon steel sheet iron loss meterThe design principle fully refers to the GB/T3655-2000 principle to ensure accurate measurement data and good repeatability, which is quite different from similar products on the market at present, and can accurately measure the mass loss Ps and magnetic induction Bm of silicon steel sheet at the same time. With the upper computer software, can be B- H hysteresis loop and u-H magnetic conductivity curve, as well as Ps-B loss curve measurement, under certain circumstances, basically meet the alternative square ring standard test silicon steel strip sample of the full performance parameters, is small and medium-sized silicon steel sheet use enterprises to control the incoming inspection of convenient and economical products.
1.
Silicon steel sheet iron loss meterInstrument features
The silicon steel sheet iron loss meter can be used to measure the iron loss of cold-rolled oriented, non-oriented and hot-rolled silicon steel sheets of various thicknesses;
The silicon steel sheet iron loss meter panel is directly operated, and the single machine can complete the test of the set point.
The silicon steel sheet iron loss meter adopts digital direct reading (unit: W/Kg), which is simple to test and has good repeatability;
The silicon steel sheet iron loss meter keeps the magnetic flux sinusoidal during the test, effectively eliminating the influence of harmonics;
Silicon steel sheet iron loss tester test probe to join the air magnetic flux compensation coil to ensure more accurate test results;
Loss comparison measurement of single stator piece P1.5/50 by open circuit magnetism meter
Closed-circuit measurement can lock the total loss of magnetic induction Bm test ratio.
Closed-circuit measurement can lock the excitation magnetic strength Hm test magnetic induction
The closed-circuit magnetometer truly compensates for zero flux
Accurate measurement of oriented and non-oriented silicon steel by closed-circuit magnetometer
Connection computer can complete the test of various curves
2.
Silicon steel sheet iron loss meterMeasuring principle
3
Silicon steel sheet iron loss meterPrecautions for use
1. The test data attached to the calibration standard chip is only used as a reference basis for the normal daily use of the instrument, and cannot be used as the standard value of the sample.
2. For the test of large-area silicon steel plates, it is recommended to use multi-point test and take the average value; the same is true for the same batch of materials.
3. When there are factors such as rust spots, stains, creases and unevenness on the surface of the tested sample, it will cause a large deviation of the test data.
4. For stamping pieces with complex shapes, the test data should be compared under the same setting conditions and at the same position.
5. When the effective contact area of the stamping piece is less than the sampling area of the probe and cannot be tested, the thickness setting can be corrected by the following formula: sampling area/effective area * sample thickness = thickness correction value
6. When the instrument is used, it should be away from magnetic field interference.
7. The tested sample piece should be placed on a flat table for testing to ensure effective contact between the probe and the sample piece and reduce the measurement error caused by poor contact.
The tested sample cannot be placed on a metal object, so as not to affect the test results.
1. When starting up
Iron loss meterIt should be connected to the magnetometer first.
2. The test cannot be locked. The phenomenon is that the test value has been 0 for a long time. You can refresh the test by pressing the "Reset" key for measurement. If the normal test still cannot be carried out, you can shut down and restart the machine. Can not work properly, please contact the manufacturer.
3. Pay attention to whether the frequency switch on the front panel is the same as the setting you need to avoid misjudgment.
4. The non-standard magnetism meter does not meet the requirements of the design standard for testing single-chip silicon steel, and can only be used as a reference for testing specific total loss. The randomly attached non-standard magnetism meter is corrected under the condition of testing non-oriented silicon steel material P1.5, which can be used to judge the quality of stator core single-chip materials of different materials, and is used to test single-chip standard samples, which is close to the real value of material loss. Since the non-standard magnetoconductivity meter does not really sample and test the magnetic induction Bm of the material, it does not provide the function of testing the magnetic induction. In principle, the wider the sample, the more accurate the tested mass is than the total loss.