FE-2100H Permanent Magnetic Material Automatic Measuring Device

FE-2100H permanent magnetic material magnetic measuring instrument is suitable for measuring the remanence Br, coercive force Hcb, intrinsic coercive force Hcj and maximum Magnetic energy product BHmax and other magnetic performance parameters, and display B&J-H hysteresis loop and demagnetization curve.
没有此类产品
Product Details
Parameter
Principle
Software
Reference
FE-2100H permanent magnet material automatic measuring device introduction
FE-2100H permanent magnet material magnetic measuring instrument is suitable for measuring the remanence Br, coercive force Hcb, internal magnetism of various ferrites, alnico, neodymium iron boron and samarium cobalt and other permanent magnetic materials. Magnetic performance parameters such as coercive force Hcj and maximum magnetic energy product BHmax, and display B&JH hysteresis loop and demagnetization curve.
FE-2100H permanent magnet material magnetic measuring instrument according to the national standard GBT 3217-2013 test method, using microcomputer control and A/D, D/A technology, automatically set the test magnetic field and flux range, automatically correct the integral zero drift and Hall The probe is non-linear. It can be equipped with a variety of fixed B/J coils according to the user's test needs, and can also be measured in different methods such as flat poles, shrink poles, heating poles, and tile poles according to test needs. No manual operation is required during the equipment measurement process, which is simple and quick. The software can directly draw the B&J-H demagnetization curve, the hysteresis loop and the measurement of the material curve at various temperatures. The test report can superimpose up to 4 temperature curves, which better meets the data comparison and analysis. The accuracy and repeatability of the instrument have reached the international advanced level.

 

FE-2100H permanent magnet material automatic measuring device applicable standards
● GB/T 3217-2013 << Permanent magnetic (hard magnetic) material magnetic test method >> 
● GB/T 24270-2009 << Measurement method for temperature coefficient of magnetic properties of permanent magnet materials >>
● IEC 60404-5 << Magnetic materials Part 5: Permanent magnetic (hard magnetic) material magnetic properties measurement method >>

 

FE-2100H permanent magnet material automatic measuring device hardware features
● Measurable materials: permanent magnet materials such as ferrite, aluminum nickel cobalt, neodymium iron boron, samarium cobalt, and rare earths;
● Measurable sample shapes: permanent magnet materials of cylindrical, square, circular and tile-shaped shapes;
● Measurable magnetic properties: magnetic properties such as remanence Br, coercive force Hcb, intrinsic coercive force Hcj and maximum magnetic energy product BHmax;
● Measurable temperature range: meet the measurement of the temperature curve of permanent magnet materials from normal temperature to 200℃; (optional)
● Features of electromagnet design: adopting pole-locking structure design, which changes the physical problem of traditional electromagnets, which is difficult to lock;
● Easy operation: In addition to setting permanent magnet measurement parameters and placing samples, the test process is completed at one time and the measurement time is short;
● Automatic range setting: adopts microcomputer control and A/D, D/A conversion technology to automatically set the test magnetic field and test range;
● Accuracy: automatically correct the integral drift and the nonlinearity of the Hall probe to ensure the accuracy of the measurement data;
● Optional test tooling:
(1) Different tooling can be selected according to sample test needs;
(2) Provide flat poles, shrink poles, heating poles or tile-type poles required for testing;
(3) Provide a fixed J coil that matches the sample.

 

FE-2100H permanent magnet material automatic measuring device software features
● Operating system: Win XP/7/8/10 Chinese or English operating system;
● Language display: support bilingual display in Chinese or English; 
● Operating features:
(1) The test waveform curve is displayed in real time during the measurement process, and the measurement can be stopped at any time;
(2) The software can superimpose and display up to 4 measurement result curves at the same time, and compare the differences under different conditions more intuitively;
(3) The software has built-in temperature coefficients of various materials, which can provide data reference for temperature correction;
(4) The measurement cycle is 30-120 seconds (depending on the sample type and the proficiency of the operator);
● Storage features:
(1) The measurement results are automatically saved (save according to the test number or manual intervention to save)
(2) Data such as sample number, size, and parameters can be automatically saved to reduce manual errors;
(3) The software has a print preview function, and can also directly print measurement results and test curves;
(4) The measurement results and test curves can be exported to graphic formats such as BMP, and can also be exported as data tables, and can be used for statistical processing with software such as Excel;

 

Corresponding parameter set not found, please add it in property template of background
FE-2100H permanent magnet material automatic measuring device main technical parameters span>
 
1, technical indicators
Under constant temperature conditions, the test meets the following requirements:
Measured parameter

HcB(%)

HcJ(%)

Br(%)

BHmax(%)

Uncertainty (K=2) 1.0 1.0 1.0 2.0
Repeatability (constant temperature) ±0.3

±0.3

±0.3

±1.0

Note: The test accuracy is traceable to samples from nationally qualified laboratories.
 
2, electromagnet
Choose different electromagnets according to the size of the sample or the size of the coercive force. There are several options:

 Model

DCT-140

 DCT-200

DCT-250 

 DCT-300

 Polar

 140mm

 200mm

 250mm

 300mm

 Polar (shrink)

 60mm

 60mm

 80mm

 100mm

 Adjustable air gap

 60mm

 80mm

80mm 

 80mm

 Magnetic field

 1.8T

 2.8T

 3.0T

 3.2T

 Resistance

 ≤5Ω

 ≤3Ω

≤2.5Ω

≤2Ω

 weight

 About 200KG

 About 800KG

 About 1400KG

 About 3000KG

 
3. Electromagnet gap magnetic field range
Model DTC-140 (shrinked to 60mm)
DTC-200 (shrinked to 60mm)

DTC-250 (shrinked to 80mm)

DTC-300 (shrinked to 100mm)

20mm gap 1.8T 2.5T 2.5T 2.5T
10mm gap 2.5T 2.9T 2.9T 2.9T
5mm gap 2.7T 3.2T 3.2T 3.2T
 
4. Power parameters
Different electromagnets with different excitation power supply configurations can generate different ranges (higher) magnetic fields. There are a variety of specifications for different customers to choose from, and the output power is from 2kVA to 16kVA.
Model 140 type power supply 200 type power supply 250 type power supply 300 type power supply
Output power 2kA 4KA 5KA 16KA
Power supply

Single-phase 220V, 50Hz

Three-phase 380V, 50Hz

Three-phase 380V, 50Hz

Three-phase 380V, 50Hz

Output voltage

±100V

±120V

±150V

±200V

Output current ±20A ±40A ±60A ±80A
Stability 0.05%
Output protection Overvoltage, overcurrent, short circuit protection
Work environment 0-40℃, 40-80% RH, no condensation
Weight 80kG
 
5, B&H instrument technical parameters
FE-210B1 flux meter (automatic range corresponding to H test)
FE-210BMagnetic flux meter is designed with the principle of RC electronic integrator, and its main performance indicators are as follows:
● Display  Display: 4 1/2 LED meter head display;
● Auto range: 2.5/5.0/10/20 mWb; each range corresponds to analog output: 0~±5V, error: ±0.5%; resolution: 1µWb;
● The magnetic flux measurement process automatically calculates the gear position according to the type of input material and the geometric size of the material;
● Drift  shift: 1µWb/minute, drift adjustment is manual;
● Resistance  Resistance: 10kΩ;
FE-210H Gauss meter (automatic range corresponding)
FE-210HGauss meter is designed using the Hall effect principle, and its main performance indicators are as follows:
● Display  Display: 4 1/2 LED meter head display;
● Auto range: 0.5, 1, 1.5, 3.0T or 0.5/1.0/2.0/4.0 T, each gear corresponds to analog output: 0~±5V, error: 0.5%, resolution: 1mT;
● The linearity is better than: ± 0.2% (computer correction); Hall plate temperature coefficient: 0.06%/℃;
● Imported Hall chip probe (linear correction range: 0~2.5T), software calibration is better than 0.3%.
● The movement of the Hall probe up and down can ensure the levelness of the Hall plate.
FE-210B2 Fluxmeter (automatic range corresponding)
FE-210BMagnetic flux meter is designed with the principle of RC electronic integrator, and its main performance indicators are as follows:
● Display  Display: 4 1/2 LED meter head display;
● Auto range: 2.5/5.0/10/20 mWb; each range corresponds to analog output: 0~±5V, error: ±0.5%;     resolution: 1µWb;
● The magnetic flux measurement process automatically calculates the gear position according to the type of input material and the geometric size of the material;
● Drift  shift: 1µWb/minute, drift adjustment is manual;
● Resistance  Resistance: 10kΩ;
Remarks: For general permanent magnets, only a Gauss meter with a B2 flux meter can meet the measurement needs, but for some high coercivity rare earth permanent magnets, the Gauss meter cannot meet the measurement needs, and double integral measurement must be used. (That is, you need to configure the B1 flux meter for measurement).
 
6. Data acquisition card
● Adopt AD to sample the analog output H and B of Gauss meter and flux meter respectively;
● Control the excitation waveform and excitation voltage amplitude through software, and control the switching of the excitation direction of the power supply at the same time;
● Resolution and linearity: 16Bit±1/2LSB; sampling frequency: 110kHz;
● Excitation and test time: 30 to 120 seconds (users can choose according to their operating proficiency);
● Conversion time between B and H: < 25μs, the synchronization effect can be ignored, the software determines Ba and Ha.
 
7, FE-200C heating power supply (non-standard configuration)
● Heating power input: 220V/5A;
● Heating power output power: 400VA;
● Control range: 50~220℃
● Heating rod: 100W/piece;
● Test probe: /pc;
● Temperature uncertainty: ±2℃, resolution 1℃
● Other non-standard small accessories (all need to be customized)
 
8. Basic configuration reference
Project Component Description
1 Excitation power Power: 2~16KVA
2 Electromagnet Diameter of pole: 140-300mm
3 Gauss meter and flux meter Gauss measurement range: 0-3T or 0-4T; magnetic flux measurement range: 0-20mwb;
4 Fixed J coil Customized according to sample size
5 Standard sample ALNiCo
6 Data Acquisition Card

12 bit or 16 bit (built-in excitation power supply)

7 Computer Industrial computer or Lenovo computer
8 Measurement software FE-2100 Magnetic Material Automatic Measuring System V1.0
9 Printer HP Laser Printer
10 Manual, certificate and warranty card Electronic file manual
Remarks: The above configuration is for customer selection reference only, and the specific configuration is subject to the contract.
 
9. Options/Accessories
● J coil: directly measure the magnetic polarization intensity without winding    
● Ordinary pure iron electrode tip: cooperate with electromagnet to produce uniform magnetic field
● Shrinking iron-cobalt alloy pole head: can generate higher excitation magnetic field, suitable for high coercivity material measurement
● Heating pole: used to measure the high temperature performance of permanent magnet materials, need to be equipped with high temperature resistant J coil
● Tile type fixture: used for magnetic tile measurement
● Temperature control power supply: 50~220℃ controllable, used with heating poles
 
The computer sends an excitation current signal to the FE-2100H excitation power supply through the data acquisition card, and at the same time controls the measurement process. The commonly used excitation signal is a full-period waveform similar to a triangular wave. Under the action of the excitation current, the electromagnet produces a changing magnetization field. At the same time, the FE-210H Gauss meter sends the magnetic field strength signal, and the FE-210B flux meter sends the magnetic flux signal to the dual-channel A/D conversion of the data acquisition card. In the device, the A/D converter converts the dual-channel voltage signal from analog to digital and sends it to the computer. After the computer obtains the whole cycle measurement signal, it uses software technology to perform various calculations and processing, and finally gives the measurement results and various graphics. The whole test process is about 60-120 seconds.
 
1. FE-2100H permanent magnet material automatic test system
1. Testing ability: meet AlNiCo, NdFeB, Ferrite, SmCo and other permanent magnet materials;
2. A set of software including Gauss meter, flux meter, data acquisition card and FE-2100 magnetic material automatic test system.
3. Highlight technical advantages
3-1. Improvement of fluxmeter drift control technology: drift control 1uWb/min, reducing errors caused by correction to ensure repeatability;
3-2. The test accuracy of Hcj and Hk is improved, and the test NdFeB under saturation magnetization has excellent repeatability;
3-3. The electromagnet has small displacement in high field, convenient to lock the pole, and flexible adjustment of the electromagnet pole support up and down;
3-4. Test software (Chinese and English): written in C#, the operating platform is compatible with Windows 2000/XP, and meets Win7;
3-5. Traceability accuracy: Permagraph C (Magnet-Physik Dr. Steingroever Gmbh);
3-6. Test repeatability (tested for all permanent magnetic materials): Br, Hcb, Hcj are better than ±0.2%, BHmax is better than ±0.5%, in Ningbo Institute of Metrology, the German Magnetic Physics Permagraph C, China Metrology The NIM-2000H of the Engineering Department of the Academy of Sciences has shown good accuracy and excellent repeatability in the accuracy and repeatability assessment experiments.
 
Second, test interface and report
1. Sampling waveform interface
2, hysteresis loop interface
3. Demagnetization curve
4. NdFeB demagnetization curve cluster
5. Test report printing
 Sample size (cylinder as an example)

Electromagnet-pole (diameter of pole face), maximum field (5mm gap)

  φ3~80mm, H3~50mm   Pure iron DCT-140 (110), maximum field 2.4T
  General permanent magnetφ3~60mm, H3~50mm

  Rare earth permanent magnetφ5~20mm, H5~15mm

  Pure iron DCT-140 (90), maximum field 2.5T

  Iron-Cobalt DCT-140B (60), the maximum field is 2.7T

  General permanent magnetφ3~120mm, H3~50mm

  Rare earth permanent magnetφ5~25mm, H5~15mm

  Pure iron DCT-200 (150), maximum field 3.0T

  Iron-Cobalt DCT-200B (60), the maximum field is 3.2T

  General permanent magnetφ3~120mm, H3~50mm

  Rare earth permanent magnetφ5~30mm, H5~15mm

   Iron-Cobalt DCT-250B (80), the maximum field is 3.2T

  General permanent magnetφ3~120mm, H3~50mm

  Rare earth permanent magnetφ5~35mm, H5~15mm

  Iron-Cobalt DCT-300B (100), the maximum field is 3.2T

The above selection is for reference only. The selection of specific equipment models and functions needs to be based on the type, brand and size of the tested sample. For details, please contact the company's technical staff for selection.

暂未实现,敬请期待
暂未实现,敬请期待
Previous article
Next article

Product Center

Related products

Key words
Key words
Key words
Key words

Relevant information

No content information display available
Please add data record on website background.