FE-2100SD-3 Soft Magnetic DC Measuring Instrument


The FE-2100SD-3 soft magnetic material DC magnetic performance test system is mainly used to measure various soft ferrite, pure iron, silicon steel, permalloy, amorphous and nanocrystalline ring samples or open-circuit bars under DC static conditions. The basic magnetization curve and hysteresis loop of the strip-shaped soft magnetic material can accurately measure the static magnetic characteristic parameters such as the initial permeability μi, the maximum permeability μm, the saturation magnetic induction intensity Bs, the remanence Br, and the coercive force Hc.



  • Product Details
  • Parameter
  • Principle
  • Software
  • Reference
    • Commodity name: FE-2100SD-3 Soft Magnetic DC Measuring Instrument
    • Commodity ID: 1303290182484250624

    The FE-2100SD-3 soft magnetic material DC magnetic performance test system is mainly used to measure various soft ferrite, pure iron, silicon steel, permalloy, amorphous and nanocrystalline ring samples or open-circuit bars under DC static conditions. The basic magnetization curve and hysteresis loop of the strip-shaped soft magnetic material can accurately measure the static magnetic characteristic parameters such as the initial permeability μi, the maximum permeability μm, the saturation magnetic induction intensity Bs, the remanence Br, and the coercive force Hc.

    Introduction

    The FE-2100SD-3 soft magnetic material DC magnetic performance test system is mainly used to measure various soft ferrite, pure iron, silicon steel, permalloy, amorphous and nanocrystalline ring samples or open-circuit bars under DC static conditions. The basic magnetization curve and hysteresis loop of the strip-shaped soft magnetic material can accurately measure the static magnetic characteristic parameters such as the initial permeability μi, the maximum permeability μm, the saturation magnetic induction intensity Bs, the remanence Br, and the coercive force Hc. (Open circuit measurement only measures Bs and Hc, of which only magnetic permeability is tested for non-magnetic materials)


    Application
      
    Automatically complete the demagnetization of closed-circuit samples, the demagnetization method is AC low-frequency demagnetization, the frequency is not higher than 10Hz, to ensure the repeatability of the initial permeability of the metallic soft magnetic material; the basic magnetization curve and permeability of the soft magnetic material of the closed-circuit sample are tested by the impact method Rate curve; the impact method is used to test the static hysteresis loop and related parameters of the soft magnetic material; the solenoid flux meter method is used to test the DC magnetic properties of the cylindrical ferromagnetic rod, and the relationship between the magnetization and the external magnetic field is implemented Demagnetization correction to obtain the relationship curve of the true magnetization intensity of the material to the magnetic field; a large number of tests are carried out on round bar semi-hard and semi-soft ferromagnetic samples with B type permeance meter; scanning method is allowed in open circuit solenoids and permeance meters use.

    Standards

    GB/T 13012-2008 Measurement method for DC magnetic properties of soft magnetic materials

    GJB 937-90 Method for measuring magnetic permeability of weak magnetic materials

    SJ/T 10281-91 Calculation of Effective 
    Parameters of Magnetic Parts
    IEC 60404-4 DC magnetic properties measurement method for soft magnetic materials

    IEC 404-7 Method for measuring the coercivity of open-circuit magnetic materials

    Key words:
    • Soft magnetic measurement
    • Soft magnetic DC measuring instrument
    • Soft magnetic material measurement
    • Magnetization curve
  • Excitation power supply

    Output Power 350VA 500VA 1000VA
    Output range ±35V/10A ±50V/10A ±50V/20A
    Output current 0.01A,0.1A,1A,10A 0.02A,0.2A,2A,20A
    Power supply:220V±10%,50Hz
    Stability: 0.05%
    Resolution: 1uA/2uA
    Accuracy: 0.1%
    数据采集控制模块

    AD part

    Input voltage: ±5 volts.
    AD converter: 16-bit AD, speed 2000-20000 channels/sec.
    Channel input impedance: 1 Mohm.
    System accuracy: ±0.1% FSR.
    Noise (peak value): less than ±1LSB (typical, 500 sampling points)
    Typical AD conversion speed:
    USB1.1 interface: 200 calls/sec. 1-16 channels can be collected per call
    DA part
     
    Two-way 16-bit DA converter.
    Output voltage: ±5 volts.
    Output voltage accuracy: ±0.1%, resolution 12 bits.
    Output voltage settling time: less than 100 microseconds.
    DA output speed: 2000-4000 times/sec (USB2.0 interface)
    Electronic integrator
    Power requirements: ±12V.
     
    Magnetic flux range: 0.25~0.5~1~2~2.5~5~10~20mwb.
     
    Resolution: 0.0001mWb.
     
    Accuracy: 0.5%.
     
    Drift: 0.001mWb/min

     

  • Instrument model

    FE-2100SD

    FE-2100SD1

    FE-2100SD3

    Instrument picture

    Ring sample

    Open circuit measurement

    Solenoid

    ×

    B-type permeability meter

    ×

    ×

    Test range

    Ring sampleui,ur,Br,Hc,Hs,Bs

    Round stick:ui,ur,Br,Hc,Hs,Bs

    Bar, stick:Br,Hc,Hs,Bs

    Sample requirements

    According toGBT13012-2008 specification ring sample requires inner and outer diameter ratio≥0.8

    Suitable for measuring ferromagnetic cylinders with aspect ratio greater than30

    Suitable for testing ferromagnetic material strips and rods. The length of the test sample must be greater than100mm.

    Test curve graph

     