FE-30SST silicon steel sheet iron loss measuring instrument


FE-30SST LOSSMETER adopts a small-scale permeance meter design (30SST) and refers to the internationally universal permeameter test standards. Through the transmission of the value traceable to the outer circle sample, the effective magnetic circuit length is determined, and complete air magnetic flux compensation is performed to meet It is very convenient and accurate to directly measure a single piece of sample. FE-30SST LOSSMETER is quite different from similar products on the market. The single machine is equipped with a calibrated permeance meter to meet the measurement of oriented and non-oriented monolithic silicon steel.



  • Product Details
  • Parameter
  • Software
    • Commodity name: FE-30SST silicon steel sheet iron loss measuring instrument
    • Commodity ID: 1303290197592133632

    FE-30SST LOSSMETER adopts a small-scale permeance meter design (30SST) and refers to the internationally universal permeameter test standards. Through the transmission of the value traceable to the outer circle sample, the effective magnetic circuit length is determined, and complete air magnetic flux compensation is performed to meet It is very convenient and accurate to directly measure a single piece of sample. FE-30SST LOSSMETER is quite different from similar products on the market. The single machine is equipped with a calibrated permeance meter to meet the measurement of oriented and non-oriented monolithic silicon steel.

    Introduction of silicon steel sheet iron loss tester
    For a long time, silicon steel sheet iron loss has been a concern of our material manufacturers, motor manufacturers, and transformer manufacturers. According to my country’s industry standards (GB/3655-2008), the use of Epstein square circle or standard Permeability meter (500SST/1000SST) is used as the basis for inspection and evaluation of silicon steel sheet iron loss. This method is an international IEC standard and national The standard recommended method of use.
    The FE-30SST LOSSMETER silicon steel sheet Iron Loss Tester adopts a small permeance meter design (30SST) and refers to the international general permeameter test standard, and confirms the effective magnetic circuit through the traceability of the measured value of the opposite coil sample Length, and complete air magnetic flux compensation, to meet the direct measurement of single-piece samples, the test is very convenient and more accurate.
    FE-30SST LOSSMETER silicon steel sheet Iron Loss Tester is quite different from similar products on the market today. The single machine is equipped with a calibrated permeance meter to meet the measurement of oriented and non-oriented monolithic silicon steel, and to more accurately measure the mass ratio total loss Ps and magnetic induction Bm of the silicon steel sheet. It can be equipped with host computer software, which can performBH hysteresis loop, uH magnetic The measurement of the conductivity curve and the Ps-B loss curve, under certain circumstances, meets the full performance parameters of the standard test silicon steel strip for the replacement of the square circle. Small and medium-sized silicon steel sheets are used by enterprises or universities with few samples, but they need to be tested on the samples. Convenient and economical small-scale test equipment under the circumstances.

     

    Characteristics of silicon steel sheet iron loss tester
    ● The panel is directly operated, and the single machine can complete the test of the set point
    ● Comparative measurement of the loss of the open-circuit permeance meter to the single stator piece P1.5/50
    ● Closed circuit measurement can lock the total loss of magnetic induction Bm test ratio
    ● Closed circuit measurement can lock the excitation magnetic strength Hm to test the magnetic induction
    ● The closed-circuit permeance meter truly achieves zero magnetic flux compensation
    ● Accurate measurement of oriented and non-oriented silicon steel by closed-circuit permeance meter
    ● It can be connected to the computer to complete the test of various curves

     

    Key words:
    • steel
    • silicon
    • loss
    • magnetic
    • test
    • BH hysteresis loop
  • Instrument parameters of silicon steel sheet iron loss tester
    First name Name Refer to  Number
    Power input 220V ±20V/0.5A,50HZ
    Output power 15V/1.5A
    Test frequency 50Hz, 60Hz free choice
    Test mode Two modes: lock B to measure H value and lock H to measure B value
    Bm setting range 500mT~1900mT
    Hm setting range 100A/m~10000A/m
    Accuracy PS accuracy is 2.0%, Bm accuracy is 1.0%
    Volume 280mm*90mm*300mm
    Sample requirements Length: more than 100mm, width: 20mm~30mm
    Sample requirements Thickness: 0.1mm to 0.65mm
    Sample requirements The samples are required to be flat, and strictly speaking, the samples need to be annealed to be traceable to the national standard.
     
    Silicon steel sheet iron loss testerBasic configuration
    Sequence number Product name Type  号 Number   Quantity
    1 Test host FE-30SST 1
    2 Monolithic Silicon Steel Permeability Meter FE-30SST 1
    3 Connection line Five-core aviation connector 1
    4 Manual CD FE-30SST 1
    5 Two silicon steel samples (manufacturer testing) Grade: 470 1
    Note: The above standard configuration is for reference only, and the actual product list is subject to the signed contract.
     
    Silicon steel sheet iron loss testerOptional Configuration
    Sequence number Product name Type  号
    1 Measurement software FE-2100
    2 Test computer Lenovo brand computer (subject to contract)
    3 Printer HP laser black and white printer (according to the contract)
    4 Open measuring probe There are certain requirements for sample width and length (20mmX40mm)
    5 Third-party approved silicon steel samples (16 pieces) Issued by Zhejiang Institute of Metrology
    Note: Users can order according to their own needs, and can also order non-standard parts according to special requirements.
  • Software function of silicon steel sheet iron loss tester
    ● The software system supports mainstream operating systems such as WinXP/NT/7
    ● The test process is fully automatic, no manual operation is required (only the sample is required)
    ● The magnetic induction (B) can be locked to measure the magnetic field (H) value, and the magnetic field (H) can also be locked to measure the magnetic induction (B) value
    ● Real-time waveform acquisition during the test, and the operation can be stopped at any time
    ● The software adopts a database file format, supports direct printing and exports the results to an Excel table
    ● You can directly save, delete, and clear data.
    ● Multi-point measurement is possible, and data can be clustered.
    ● Through multi-point clustering, the hysteresis loop, magnetization curve, permeability curve and loss curve can be displayed.
    ● Supports various types of printers, and can print test reports directly
    ● The test report can be generated in JPG, Bmp file format, and the report file can be sent via Email

     

    Silicon steel sheet iron loss test interface

    Sampling waveform diagram with test software
    Silicon steel sheet magnetization curve test chart
    Hysteresis loop (cluster) analysis interface
    Silicon steel sheet loss curve interface