FE-2100SD DC measuring device for soft magnetic materials

The FE-2100SD soft magnetic material DC measurement system is a comprehensive soft magnetic measurement equipment that can automatically measure the basic magnetization curve and basic hysteresis loop of soft magnetic materials under DC conditions. It conforms to GB/T13012-2008 and other soft magnetic materials. The DC measurement standard can accurately measure the static magnetic characteristic parameters such as initial permeability μi, maximum permeability μm, saturation magnetic flux density Bs, remanence Br, coercive force Hc and hysteresis loss Pu.
没有此类产品
Product Details
Parameter
Principle
Software
Reference
Product introduction
FE-2100SD DC measurement system for soft magnetic materials, which automatically measures the basic magnetization curve and hysteresis loop of soft magnetic materials under static (DC) conditions, and accurately measures the initial permeability μi, maximum permeability μm , Saturation magnetic induction intensity Bs, remanence Br, coercive force Hc and hysteresis loss Pu and other static magnetic characteristic parameters.
According to the measurement principle of the impact method, the test system adopts computer control technology and A/D, D/A combination, replaces the traditional impact galvanometer with an electronic integrator, realizes the simulation impact measurement under the control of the computer, and completely eliminates the classic impact In the method, the non-instantaneous error caused by the use of the impact galvanometer improves the measurement accuracy and provides a reliable basis for studying the mechanism of the material magnetization process.
 
FE-2100SD soft magnetic material DC measuring device applicable standards
◉ GB/T 13012-2008 Measurement method of DC magnetic properties of soft magnetic materials
◉ GJB 937-90 weak magnetic material permeability Measurement method
◉ GJB 937-90 weak magnetic material permeability Measurement method
◉ SJ/T 10281-91 Magnetic parts effective parameters Calculation
◉ IEC 60404-4 DC magnetic properties of soft magnetic materials Measurement method
◉ IEC 404-7 open circuit magnetic material coercivity Measurement method
 
Hardware characteristics of FE-2100SD soft magnetic material DC measurement device
◉  automatically complete the demagnetization of closed-circuit samples, the demagnetization method is AC low-frequency demagnetization, the frequency is not higher than 10Hz, to ensure the repeatability of the initial permeability of the metallic soft magnetic material;
◉ Using the impact method to test the basic Magnetization curve and permeability curve;
◉  Use impact method to test the static hysteresis of soft magnetic materials Return line and related parameters;
◉  allows the use of scanning method in open solenoid and love Used in Perstein square circle;
 
FE-2100SD soft magnetic material DC measurement device software features
◉  automatically complete the data calculation and automatically correct the span coefficient. The system supports mainstream operating systems such as Winxp/NT/7/8.
◉  show I(t), B(t) samples Waveform and B(H) hysteresis loop.
◉  can display B(H) hysteresis loop cluster , B(H) magnetization curve, μ(H) permeability curve.
◉  The instrument is easy to operate and can be interrupted by human intervention.
◉  has powerful file management functions. It has the functions of automatically saving data, deleting data, and clearing all data.
◉  The data file contains complete sampling data and sample parameters And instrument parameters, the file adopts text format, which can be easily input into other software.
◉ The test report contains complete graphs, test results, Test conditions and sample parameters.
◉  The file system adopts database format, which can be directly printed or output The test results are transferred to an Excel table.
◉  can be in Chinese and English format, which is convenient for adding users Logo and company name.
◉  has a print preview function for easy adjustment and testing The size and margins of the report.
◉  can print the test report directly, or generate the test report JPG picture file.
◉  can send JPG image files directly via E-mail, Or save the JPG picture file to disk.
◉  supports various types of printers, test reports and printers The paper matches exactly.
 
Corresponding parameter set not found, please add it in property template of background
Main technical parameters of FE-2100SD soft magnetic material DC measuring device
 
1. Excitation current source
There are four excitation current sources for users to choose from: 350VA, 500VA, 1000VA, and 2000VA.
350VA
◉ Power supply requirements: 220V± 10%, 5A, 50Hz.
◉ Output range:±35V/10A.
◉ Current range: 0.01A, 0.1A, 1A, 10A.
◉ Stability: 0.05%.
◉ Resolution: 0.1mA.
◉ Accuracy: 0.1%.
 
500VA
◉ Power requirements: 220V±10%, 5A, 50Hz.
◉ Output range:±50V/10A.
◉ Current range: 0.01A, 0.1A, 1A, 10A.
◉ Stability: 0.05%.
◉ Resolution: 0.1mA.
◉ Accuracy: 0.1%.
 
1000VA
◉ Power supply requirements: 220V±10%, 10A, 50Hz.
◉ Output range:±50V/20A.
◉ Stability: 0.05%.
◉ Resolution: 0.2mA.
◉ Accuracy: 0.1%.
 
2000VA
◉ Power supply requirements: 380V±10%, 20A, 50Hz.
◉ Output range:±50V/40A.
◉ Current range: 40mA, 400mA, 4000mA, 40A.
◉ Stability: 0.05%.
◉ Resolution: 0.4mA.
◉ Accuracy: 0.1%.
 
 
2. Data acquisition module
AD section
◉ Input voltage:±5 volts.
◉ AD converter: 16-bit AD, speed 2000-20000 channels/sec.
◉ Channel input impedance: 1 Mohm.
◉ System accuracy:±0.1% FSR.
◉ Resolution: 16 bits.
◉ Noise (peak): less than ±1LSB (typical, 500 sampling points)
◉ Typical AD conversion speed:
◉ USB1.1 interface: 200 calls/sec. Each call can collect 1-16 channels.
◉ USB2.0 interface (scan mode): single channel 2000-4000 channels/sec, 16 channel acquisition time 1ms (equivalent to AD conversion speed 16KHz).
◉ USB2.0 interface (single-channel average mode): 100 times average, sampling period 2.5ms, equivalent AD conversion speed: 40KHz.
 
DA part
◉ Two-way 16-bit DA converter.
◉ Output voltage:±5 volts.
◉ Output voltage accuracy: ±0.1%, resolution 16 bits.
◉ Setup time of output voltage: less than 100 microseconds.
◉ DA output speed: 2000-4000 times/sec (USB2.0 interface)
 
3. Electronic integrator
◉ Power requirements:±12V.
◉ Magnetic flux range: 0.25~0.5~1~2~2.5~5~10~20mwb.
◉ Resolution: 0.0001mWb.
◉ Accuracy: 0.5%.
◉ Drift: 0.001mWb/min.
 
4. Reference configuration of FE-2100SD soft magnetic material DC measuring device
◉ FE-2100 Magnetic Material Automatic Measuring System V1.0          one set
◉ FE-2100SD soft magnetic material magnetic material measuring host         one set                            < span style="font-size: 12pt;">◉ Lenovo OR Advantech Industrial Computer                 One set (optional)
◉ HP Laser Printer                     one (optional)
◉ Standard sample                         one (optional)
◉ Standard cabinet and workbench                    one set (optional)
◉ CD and manual                       a set
 
5. Uncertainty of FE-2100SD soft magnetic material DC measurement device
Measured parameter Bs (%) Hs (%)

μi(%)

Br(%) Hc (%)
Uncertainty 1.0 1.0 5.0 1.0 2.0
Repeatability (constant temperature) ±0.3

±0.3

±2.0

±0.3

±0.3

 

The FE-2100SD soft magnetic material DC measurement system is a comprehensive soft magnetic measurement equipment that can automatically measure the basic magnetization curve and basic hysteresis loop of soft magnetic materials under DC conditions. It conforms to GB/T13012-2008 and other soft magnetic materials. The DC measurement standard can accurately measure the static magnetic characteristic parameters such as initial permeability μi, maximum permeability μm, saturation magnetic induction Bs, remanence Br, coercive force Hc and hysteresis loss Pu.
In the measurement of the DC characteristics of soft magnetic materials in the magnetization curve, the most difficult to test the accurate amount is the initial permeability. The test error can reach tens of percent with the impact galvanometer or the scanning method. Most domestic products use simulation The impact method effectively avoids the influence of fluxmeter drift. The FE-2100SD soft magnetic material DC measurement system inherits the technology of the simulated impact method, improves the impact waveform and signal, and at the same time continuously improves the integrator drift control technology, and optimizes the control of the test point, and through the excellent performance of dB/ The dt control technology, high-resolution current control technology, and magnetic flux sampling technology make the entire test extremely well controlled. It has obvious advantages over domestic and foreign counterparts in the testing of micro-components, low-Br materials and iron powder core materials.
In 2016, according to the vision of Mr. Qu Qingchang, one of the founders of the DC magnetic performance measurement device for soft magnetic materials in China, and a more in-depth study on the hysteresis loop measurement method of soft magnetic materials. Up to half a year of experimental comparison and update of the test equipment method found the defects of the current magnetic measurement equipment, and completed the following improvements.
1. There is a large uncertainty in the measurement of soft magnetic materials under DC conditions using the scanning method;
2. The definition of the basic hysteresis loop is proposed, that is, the point-to-point simulation impact method is used to measure the hysteresis loop of the material to ensure the uniqueness of the hysteresis loop and the relevant magnetic characteristic parameters Br/Hc/Bs/Hs;
3. Different materials use different impact times and impact waveforms. Only when the impact frequency reaches a certain requirement (slow enough), can the uniqueness of the hysteresis loop and related parameters be guaranteed.
Participated in the Coil Fair in Berlin, Germany in 2014, and was affirmed by international companies and foreign counterparts, and has begun to provide this device to well-known international companies, and has been exported to scientific research institutes and companies in the United States, Germany, Israel, Russia and other countries.
 
Software acquisition waveform interface :
 
software magnetization curve interface :
 
Software hysteresis Line test interface
 
Software magnetization curve printing Performance

Instrument model

FE-2100SD

FE-2100SD1

FE-2100SD3

Instrument picture

Ring sample

Open circuit measurement

Solenoid

×

B-type permeability meter

×

×

Test range

Ring sampleui,ur,Br,Hc,Hs,Bs

Round stick:ui,ur,Br,Hc,Hs,Bs

Bar, stick:Br,Hc,Hs,Bs

Sample requirements

According toGBT13012-2008 specification ring sample requires inner and outer diameter ratio≥0.8

Suitable for measuring ferromagnetic cylinders with aspect ratio greater than30

Suitable for testing ferromagnetic material strips and rods. The length of the test sample must be greater than100mm.

Test curve graph

 

暂未实现,敬请期待
暂未实现,敬请期待
Previous article
Next article

Product Center

Related products

Key words
Key words
Key words
Key words

Relevant information

No content information display available
Please add data record on website background.